제품 정보
Iwatsu · PMK 제품
| Item | HV-P30A | |
| Maximum input voltage | DC:60kV single pulse:80kV (standard pulse wave:tip length = 1µs/tail length = 40µs) |
|
| Maximum input voltage | DC:30kV single pulse:40kV (standard pulse wave:tip length = 1µs/tail length = 40µs) |
|
| Frequency bandwidth | DC~50MHz (-3dB) | |
| Rise time | 7ns or less | |
| Input RC | R | R:100MΩ±5% |
| C | 7pF or less | |
| Attenuation | 1000:1 ±5% | |
| Waveform distortion | ±5% (10℃ to 35℃) * However, except for ringing and overshooting | |
| Cable length | Approx. 3m | |
| Connector type | BNC type | |
| Connectable Oscilloscope input RC | R | 1MΩ |
| C | 10pF~40pF | |
| Size | See exterior view | |
| Operating temperature/Storage temperature | -10℃~+50℃ | |
| Weight | Probe main unit:Approx. 1.3kg overall:Approx. 2.5kg (including accessories) |
|
| Accessories | Output cable (including matching box), clamping bracket, tip, earth lead, adjustment screwdriver, storage bag | |
| Option | High pressure probe stand (made-to-order product) SK-301 | |